From LipidomicsWiki
Contents |
Instrument
- Type: Pharmacia Smart System® including fraction collector
- Running buffer: Dulbecco's PBS containing 1mM EDTA
- Flow: constant flow of 40μL/min
Column
- Type: Superose 6 PC 3.2/30 column (GE Healthcare)
- Temperature: room temperature
Fractionation
- Injection volume: 50 µL human serum
- Fractionation start: 18min
- Fraction volume: 80µl
- Number of fractions: 20
Example
Analysis of cholesterol and triglyceride content of the fractions of a human normolipidemic serum.